Type: Trabalho de Conclusão de Curso
Title: Avaliação da radiopacidade dos cimentos MTA angelus® branco, MTA repair HP e biodentine por meio de radiografia digital
Author: Fiche, Gabriela El-Corab
First Advisor: Resende, Leandro Marques de
Co-Advisor: Leite, Anamaria Pessôa Pereira
Referee Member: Resende, Leandro Marques de
Referee Member: Leite, Anamaria Pessoa Pereira
Referee Member: Devito, Karina Lopes
Resumo: O objetivo deste estudo foi avaliar as radiopacidades dos cimentos MTA Angelus branco (MTA B), MTA Repair HP (MTA HP) e Biodentine (BD) e compará-las com as das estruturas dentais humanas, por meio de radiografia digital. Para tal, foram confeccionados cinco corpos de prova, com 4 mm de diâmetro e 2 mm de altura, para cada cimento estudado. Cada amostra foi radiografada, pelo método digital, juntamente com uma fatia de 2 mm de espessura de um dente molar inferior humano hígido e uma escala de densidade de alumínio (Al), variando de 2 a 16 mm, com incrementos de 2 mm por degrau. O tempo de exposição bem como a distância foco-filme foram padronizados para todas as exposições. Para cada conjunto, três imagens foram obtidas e avaliadas quanto às densidades por meio da ferramenta histograma do software ImageJ. Em cada estrutura avaliada: esmalte, dentina, MTA B, MTA HP, BD e os degraus da escala de densidade, foi usada uma região de interesse pré-definida e as densidades foram apresentadas utilizando-se médias, desvios-padrões e medianas. Para avaliar a normalidade dos dados e comparar a radiopacidade dos três cimentos testados, foram aplicados os testes de Shapiro-Wilk e o de Kruskal-Wallis, respectivamente. Foi utilizado o programa Statistical Package for the Social Sciences (SPSS) com nível de significância de 5% (p ≤ 0,05). Após a análise das diversas estruturas, foi observado para o esmalte e a dentina uma radiopacidade equivalente a 2 mm de Al para o primeiro e 1 mm de Al para o segundo. O teste de Kruskal-Wallis não mostrou diferença significativa entre as densidades dos cimentos MTA B e MTA HP testados, uma vez que ambos apresentaram radiopacidade equivalente a 3 mm de Al. No entanto, o cimento BD apresentou uma radiopacidade significantemente inferior aos demais materiais testados, equivalente a 2 mm de Al. Assim, sugere-se que tanto o MTA B quanto o MTA HP apresentaram radiopacidades em conformidade com as normas da International Organization for Standardization 6876/2001 e American National Standard Institute número 57, ao contrário do BD, que demonstrou radiodensidade inferior e não compatível com a regulamentação.
Abstract: The aim of this study was to evaluate the radiopacities of MTA Angelus white (MTA B), MTA Repair HP (MTA HP) and Biodentine (BD) cements and compare them with those of human dental structures, by digital radiography. For this purpose, five specimens were made, with 4 mm in diameter and 2 mm in height, for each studied cement. Each sample was radiographed by the digital method, along with a 2 mm thick slice of a healthy lower human molar tooth and an aluminum density scale (Al), ranging from 2 to 16 mm, in increments of 2 mm per step. The exposure time as well as the focus-film distance were standardized for all exposures. For each set, three images were obtained and evaluated for densities using the histogram tool of the ImageJ software. In each evaluated structure: enamel, dentin, MTA B, MTA HP, BD and the steps of the density scale, a region of pre-defined interest was used and the densities were presented using means, standard deviations and medians. To assess the normality of the data and compare the radiopacity of the three cements tested, the Shapiro-Wilk and Kruskal-Wallis tests were applied, respectively. The Statistical Package for the Social Sciences (SPSS) program was used with a significance level of 5% (p ≤ 0.05). After the analysis of the different structures, a radiopacity equivalent to 2 mm Al for the first and 1 mm Al for the second was observed for the enamel and dentin. The Kruskal-Wallis test did not show any significant difference between the densities of the MTA B and MTA HP cements tested, since both had radiopacity equivalent to 3 mm of Al. However, the BD cement showed a radiopacity significantly lower than the other tested materials , equivalent to 2 mm of Al. Thus, it is suggested that both MTA B and MTA HP presented radiopacities in accordance with the standards of the International Organization for Standardization 6876/2001 and American National Standard Institute number 57, in contrast to BD, which demonstrated lower radio-density and not compatible with the regulation.
Keywords: Radiopacidade
Radiopacity
MTA Angelus® Branco
White MTA Angelus®
MTA Repair HP
HP Repair MTA
Biodentine
Biodentine
CNPq: CNPQ::CIENCIAS DA SAUDE::ODONTOLOGIA::ENDODONTIA
Language: por
Country: Brasil
Publisher: Universidade Federal de Juiz de Fora (UFJF)
Institution Initials: UFJF
Department: Faculdade de Odontologia
Access Type: Acesso Restrito
Creative Commons License: http://creativecommons.org/licenses/by/3.0/br/
URI: https://repositorio.ufjf.br/jspui/handle/ufjf/12402
Issue Date: 10-Jan-2020
Appears in Collections:Faculdade de Odontologia - Campus JF - TCC Graduação



This item is licensed under a Creative Commons License Creative Commons