Please use this identifier to cite or link to this item: https://repositorio.ufjf.br/jspui/handle/ufjf/8307
Files in This Item:
File Description SizeFormat 
Avaliação da resistência à corrosão de implantes.pdf798.11 kBAdobe PDFThumbnail
View/Open
Type: Artigo de Periódico
Title: Avaliação da resistência à corrosão de implantes em aços inoxidáveis ISO 5832-9 e F138 removidos de pacientes
Other Titles: Evaluation of the corrosion resistance of stainless steel ISO 5832-9 and F138 implants removed from patients
Author: Silva, Elison da Fonseca e
Oliveira, Luiz Fernando Cappa de
Resumo: -
Abstract: This work has compared the surfaces of two different steel samples used as orthopedical implants, classified as ASTM F138 and ISO5832-9, through optical emission spectroscopy, by means of SEM and EDS. The samples (implants) were also submitted to potentiodynamic cyclic polarization in Ringer lactate and NaCl 0.9 M L-1 solutions; ISO5832-9 sample did not show any kind of localized corrosion, but in the case of F138 steel was observed a pit localized corrosion in both solutions. In Ringer lactate solution it was observed a loss of about 63% for nickel and 26% for iron for F138 stell, compared to the initial composition.
Keywords: Stain steel
Orthopedic implant
Corrosion
CNPq: -
Language: por
Country: Brasil
Publisher: -
Institution Initials: -
Access Type: Acesso Aberto
DOI: http://dx.doi.org/10.1590/S0100-40422012000100012
URI: https://repositorio.ufjf.br/jspui/handle/ufjf/8307
Issue Date: 2012
Appears in Collections:Artigos de Periódicos