Abstract
This study aimed to develop a method for arsenic extraction from atmospheric
particulate matter samples using an ultrasonic bath, followed by quantification by
hydride generation atomic absorption spectrometry (HG-AAS). Initially, the operational
conditions were investigated to obtain the best instrumental performance, and the most
suitable analytical figures of merit were achieved at a peristaltic pump speed of 40 rpm,
an Ar carrier gas flow rate of 100 mL min-1
, a delay time of 50 s, 5.0 mol L
-1 HCl, and
1.25% (m v
-1
) NaBH₄, with pre-reduction using KI and L-ascorbic acid 2.0% (m v
-1
).
During the development of the sample preparation procedure, urban dust was used as a
test sample and EPA Method 3050B as a reference, yielding an arsenic concentration
with a relative standard deviation (RSD) below 5.0%, indicating good precision of the
method employed. In this method, recoveries at the low, medium, and high levels
ranged from 91 to 101%, and no matrix effect was detected under the conditions
evaluated. Subsequently, the efficiency of HCl, HNO3
, and H3PO4
for arsenic extraction
in combination with an ultrasonic bath was evaluated using a mixture design.
Satisfactory results were obtained using HCl, a binary HNO3
-H3PO4 mixture at a 3:1 v v
-1
ratio, and a ternary mixture at a 1:1:1 v v
-1
ratio, with recoveries above 90%. In addition
to the extracting agents, the effects of extraction time and acid concentration were also
investigated, and satisfactory results were achieved using concentrated HCl for 10 min
of extraction in an ultrasonic bath at 60 °C. To apply the developed method, it was used
for arsenic extraction in five PM2.5 samples collected in Juiz de Fora, Minas Gerais,
Brazil, using MicroVol-1100 samplers, with PM2,5 concentrations ranging from 11.9 to
64.0 µg m-3
. Daily arsenic concentrations ranged from 20.7 to 1149 ng m-3
,
demonstrating the adequate applicability of the developed ultrasonic extraction method
followed by HG-AAS quantification. Under the optimized conditions, a working range of
0.5 to 5.0 µg L
-1 yielded limits of detection (LOD) and quantification (LOQ) of 0.12 and
0.35 µg L
-1
, respectively. Thus, arsenic determination in PM₂.₅ was shown to be feasible
using less extraction time, sample mass, and reagent consumption, in addition to
generating less waste. The results also indicated the need for broader studies to assess
arsenic levels in PM2.5
in Juiz de Fora, Minas Gerais, Brazil, their sources, and
variations over different periods, given the scarcity of studies on this topic in the city.