Comparação das propriedades mecânicas e superficiais de diferentes resinas acrílicas utilizadas na confecção de placas estabilizadoras
Loading...
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Universidade Federal de Juiz de Fora (UFJF)
Abstract
Introduction: Currently, different types of acrylic resins have been used for the
fabrication of occlusal splints. These resins may be processed using different
manufacturing techniques, which may result in differences in their physicomechanical
properties. Objective: To compare the surface roughness, hardness, flexural strength,
and failure patterns of different acrylic resins used for the fabrication of occlusal
splints processed by distinct manufacturing techniques: chemically activated acrylic
resin (CAAR), heat-activated acrylic resin (HAAR), 3D-printed resin (PR), and milled
resin (MR). Methods: Fifteen bar-shaped specimens (65 × 10 × 2.5 mm; ISO 20795-
1) were fabricated for each experimental group. Specimens from the HAAR (Artigos
Odontológicos Clássico Ltda.) and CAAR (Artigos Odontológicos Clássico Ltda.)
groups were produced using flask molds according to their respective conventional
polymerization protocols. For the MR (SmileFind Dental) and PR (Smart Print) groups,
the specimens were initially designed as STL files using CAD technology with MillBox
2023 (CIMsystem, Italy) and CHITUBOX Dental version 1.2 (CBD-Tech/Chitu
Systems, China), respectively, and subsequently manufactured according to the
corresponding CAM technology for each group. After fabrication, all specimens were
polished with 600-, 1200-, and 2000-grit silicon carbide abrasive papers using a
mechanical polishing machine, followed by ultrasonic cleaning in isopropyl alcohol for
5 minutes and storage in distilled water at 37 °C for 48 hours. Surface roughness (Ra)
was measured using a contact profilometer (0.25 mm cutoff; 0.5 mm/s; n = 15), flexural
strength was determined by a three-point bending test (100 kgf; 5 mm/min; n = 12),
and Knoop hardness was evaluated using a microhardness tester (n = 3; three
indentations per specimen spaced 16 mm apart; 10 s; 50 g). Failure patterns were
analyzed using a stereomicroscope at 12× magnification. Quantitative data were
analyzed by one-way analysis of variance (ANOVA) followed by Tukey's post hoc test
(α = 0.05). Results: The printed resin exhibited significantly lower surface roughness
(CAAR: 0.17 ± 0.06 µm; HAAR: 0.14 ± 0.06 µm; MR: 0.14 ± 0.08 µm; PR: 0.11 ± 0.03
µm; p = 0.006) and higher flexural strength (CAAR: 125.10 ± 19.20 MPa; HAAR:
126.43 ± 21.41 MPa; MR: 150.58 ± 16.73 MPa; PR: 161.84 ± 13.89 MPa; p < 0.001)
than the other groups, which did not differ significantly from one another for these
properties. Knoop hardness was significantly higher for the heat-activated acrylic resin
(CAAR: 15.47 ± 1.04 HK; HAAR: 18.62 ± 2.03 HK; MR: 18.50 ± 1.60 HK; PR: 17.62 ±
0.73 HK; p < 0.001) compared with the other groups, which also showed no significant
differences among themselves. Failure pattern analysis demonstrated that all
fractures, regardless of the group, originated from the tensile stress region; however,
the PR group exhibited a more well-defined and clearly demarcated fracture pattern
than the other groups. Conclusion: The choice of material and processing technique
directly influences the final quality of occlusal splints. Milled acrylic resins appear to be
the most suitable option when superior mechanical and surface performance is
desired.