Metodologia para o controle de qualidade e segurança do leite em relação à presença de contaminantes inorgânicos

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

Universidade Federal de Juiz de Fora

DOI

Abstract

This study aimed to establish a method for the determination of inorganic contaminants in raw milk that was compatible with largescale applications, such as regulatory programs for the monitoring of contaminants in milk. For this purpose, an approach using the technique of atomic emission spectrometry (ICP-AES) with no mineralization of the samples, which was solubilized in a solution containing ethanolamine and citric acid, has been proposed. Studies were conducted in order to establish experimental conditions in which it was possible to use the proposed sample preparation and keeping the commitment to sensitivity, accuracy and the analytical performance required by legislations related to food safety. The ideal ratio of sample dilution was 1:2 (sample:solution) and the most suitable values for instrumental parameters were 1300W for the plasma radiofrequency power, 0.50L.min-1 for the flow of auxiliary gas and 0.60L.min-1 for the flow of nebulizer gas. The quantification limits obtained for Cd, Co, Cr, Cu and Pb were, respectively, 6.54, 6.57, 14.8, 14.3 and 27.4 μg.mL-1.

Description

Citation

Endorsement

Review

Supplemented By

Referenced By

Rights and licensing

Acesso Aberto