Contaminação microbiana em diferentes identificadores de instrumental cirúrgico odontológico
Loading...
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Universidade Federal de Juiz de Fora (UFJF)
DOI
Abstract
Cross-infection through dental instruments begins with the adhesion of
microorganisms to their surface, favored by the presence of indentations and
roughness that make it difficult to remove dirt. In view of this, it is believed that the
use of surgical instrument identifiers is capable of enhancing this microbial retention.
The study aimed to analyze microbial contamination in different dental instrument
identifiers. To this end, 15 n°3 scalpel handles were used as specimens, divided into
3 groups (n=5) in order to obtain three biological replicates. Each group consisted of
four instruments that had different identifiers (laser engraving, silicone ring, adhesive
tape and personalized adhesive label) and a control group, which corresponded to
the instrument without an identifier. The scalpel handles were exposed to the same
organic material for 30 minutes and were subsequently subjected to the cleaning and
disinfection process in accordance with the standards contained in technical report n°
01/09 of the Agência Nacional de Vigilância Sanitária and the autoclave sterilization
process. Microbial growth was evaluated at different times, for a period of 24 hours to
14 days, through qualitative and quantitative analyses, by observing the turbidity of
the culture medium and counting colony-forming units. Microbial contamination was
observed in all instruments, regardless of the identifier used. However, it is noted that
this contamination did not occur continuously and progressively after repeating the
cleaning and sterilization processes. In conclusion, the study demonstrated that the
use of different dental instrument identifiers is not associated with microbial
contamination, as long as cleaning and sterilization standards for these materials are
respected, highlighting the importance of interdependence between these steps for
effective removal of microorganisms.
Description
Keywords
Citation
Endorsement
Review
Supplemented By
Referenced By
Creative Commons license
Except where otherwise noted, this item's license is described as Acesso Aberto