Within the scope of this study, two low-cost electrochemical platforms were meticulously
engineered for the quantification of amoxicillin in samples of whole milk, tablets, and synthetic
urine. The paper electrode not only evinced reproducibility in measurements but also exhibited
an expansive linear working range. A conductive ink was formulated, employing base enamel,
acetone, and graphite powder, with scrupulous attention to optimizing its composition
concerning the percentage of conductive material. A noteworthy discrepancy in sensor
performance was discerned upon the mechanical exfoliation of its surface, resulting in an
approximate 30% reduction in its charge transfer resistance. Scanning electron microscopy
analyses revealed an augmentation in surface homogeneity and an enhancement in the exposure
of electroactive sites post-process. A method utilizing differential pulse voltammetry was
refined, yielding a linear sensor response within the range of 29.8 to 126.7 µmol L-1
. In real
sample tests, under an addition and recovery assay, recovery values ranging from 90% to 103%
were attained, with a detection limit of 8 µmol L-1
. Furthermore, no substantive interferences
were discerned in the analytical signal in the presence of other clinically relevant
pharmaceuticals. The visible laser-engraved electrode was meticulously constructed with
accessibility, high scalability, and reproducibility in mind. Analyses through EIS, SEM, and
RAMAN unveiled a porous surface with graphene morphologies housing abundant
electroactive sites. Notably, this platform exhibited a diminished charge transfer resistance
when juxtaposed with a commercial glassy carbon electrode. An electroanalytical method
employing differential pulse voltammetry was devised, achieving a working range of 14.9 to
126.2 µmol L-1
and a detection limit in the order of 4.9 µmol L-1
. When applied to
pharmaceutical and food samples, in an addition and recovery assay, recovery values fluctuated
between 86% and 104%. In summation, both platforms epitomize promising alternatives for
the analysis of the proposed samples, featuring micromolar order working ranges, coupled with
precision, accuracy, and auspicious scalability.