Preparação de novas fases estacionárias monolíticas para uso em eletrocromatografia capilar
Loading...
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Universidade Federal de Juiz de Fora (UFJF)
DOI
Abstract
This thesis describes the preparation of new monolithic stationary phases
(MSP) polymerized by photoinitiation through sol-gel approach in polyacrylate-coated
fused silica capillary, for application in capillary electrochromatography (CEC). CEC
has been concentrated much attention among the major separation techniques in
analytical chemistry because it combines the advantages of both high performance
liquid chromatography and capillary electrophoresis. Much of the CEC development
is due to the use of MSP, which are similarly applied to other chromatographic
techniques. Unlike polyimide-coating, widely used, the polyacrylate-coating, which is
transparent above 370 nm and visible, enables the visualization of the sol solution
within the capillary, allowing one to control the injection of sol and other solutions, in
addition to observe the in situ formation of the MSP. Furthermore, it is possible to
perform the photoinitiated polymerization without removing this polymeric coating that
protects the capillary. The main purpose of this work was to comprehend and
improve the fabrication process of MSP, for CEC applications. For this, some
changes were set in the homemade photo reactor chamber, used for the MSP
polymerization, like correction in the work range from 350 – 700 nm to 350 – 400 nm;
and installation of security devices for both operator and electric system safeties. For
better control of liquid phases injection within tubes with capillary dimensions, the
build of a high-pressure device (HPD) that provides a great precision, in addition to
the high-pressure, was essential. HPD is simpler and relatively cheaper when
compared to commercial models. The preparation of the MSP has been optimized through assistance of a 24-1 fractional factorial design, with the intention to investigate
electrochromatographic properties with different amounts of employed reagents and
ultraviolet (UV) light incidence time. The later factor did not show significance and was unconsidered, making the design possible to be reduced to a 23 complete
factorial design, which allowed analyzing the significant effects accurately. The most
influent factor was the porogen (toluene) proportion, and the best condition was
obtained using 80.0 % (v/v) of porogenic solution; 3.5 % of photoinitiator bis(2,4,6
trimetylbenzoyl)-phenylphosphine oxide (Irgacure 819); water to
metacryloxypropyltrimethoxysilane (MPTMS) molar ratio equal to 4 and 10 minutes of
UV light incidence time. The MSP morphological, spectroscopic characteristics and
porosity were evaluated through scanning electron microscopy, infrared spectroscopy
and nitrogen adsorption porosimetry, respectively. The MSP has been tested in CEC
through the separation of polycyclic aromatic hydrocarbons (naphthalene,
acenaphthene, fluorene, phenanthrene and anthracene) and alkylbenzenes
(ethylbenzene, propylbenzene, butylbenzene and hexylbenzene), which are electrically neutral compounds, after dilution in methanol (1 mmol L-1 each), using
thiourea as the flow marker. As mobile phase a mixture of ammonium acetate 16.7 mmol L-1 at pH 7.0 (60.0 %) and acetonitrile (40.0 %) was used. The applied
voltage was -20 kV, the temperature of analysis was 20 °C, the analyte injection was
-25 mbar for 5 s, and UV detection was done at 220 and 250 nm. A fast-CEC mode,
which consists to reverse the analysis direction and to introduce the analyte by
capillary short-end injection, was performed. This mode was much more fast,
repetitive and efficient than the normal one, providing in a little more than 12 minutes
over than 51400 plates per meter of column and relative standard deviations ranging
from 0.09 to 3.3 % for migration/retention time and from 0.14 to 1.6 % for relative
peak area. The separation profiles in CEC corroborate with the porosity and
morphology results.
Description
Keywords
Fase estacionária monolítica , Eletrocromatografia capilar , Eletroforese capilar , Câmara reatora fotoquímica , Fotopolimerização , Capilar de sílica revestido com poliacrilato , Monolithic stationary phase , Capillary electrochromatography , Capillary electrophoresis , Photochemical reactor chamber , Photopolymerization , Polyacrylate-coated fused silica capillary
Citation
Collections
Endorsement
Review
Supplemented By
Referenced By
Rights and licensing
Acesso Aberto