Avaliação da resistência à corrosão das limas endodônticas Protaper® Universal, Mtwo®, Reciproc® e Wave-One® imersas em solução de hipoclorito de sódio a 5,25%
Loading...
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Universidade Federal de Juiz de Fora
DOI
Abstract
This study tested the hypothesis that endodontic instruments
manufactured with the NiTi alloy heat treated M-Wire are more resistant to corrosion
in NaOCl solution. Was evaluated by potentiodynamic polarization test, the corrosion
resistance of NiTi files 14 without surface treatment (Universal ProTaper® Finishing 2
- F2, n = 7 and Mtwo® 25.07, n = 7) and 14 files NiTi with surface treatment (R25
Reciproc® 25.08, n = 7 and Primary Wave-One® 25.08, n = 7), when immersed in
irrigating solution of NaOCl 5.25%. In electronic scanning microscopy, the
micromorfologies of 28 tested files analyzed before and after the electrochemical
test. The anodic polarization curves obtained by scanning the potential Ecorr that
initiated in each file with a rate of 0.333 mV/sec (1.2 V/minute) to the Epite, being
considered passive track, the region where the values potential had density of less
than 10-5 A/cm2 current. In order to compare the averages of the groups in pairs used
to non-parametric Mann-Whitney test, using the SPSS-15 statistical software and the
significance level was p < 0.05. The anodic polarization curves show that all tested
files are liable to the corrosion solution of 5.25% NaOCl (pH = 6.5). It was observed
that, in general, only between Mtwo® (without heat treatment) and Reciproc® (heat
treated) files there was no statistically significant difference for Epite ECS and Epite-
Ecorr (p = 1.000). When comparing in pairs, the files without heat treatment
(ProTaper® Universal F2 and Mtwo®) with files with heat treatment (Reciproc® and
Wave-One®) there was no statistically significant difference (Ecorr ECS, p = 1.000;
Epite ECS, p = 0.265; Epite-Ecorr, p = 0.265). The Reciproc® file introduced more
resistant to corrosion and the ProTaper® was the file that suffered most corrosion in
the tested solution. Thus, the initial hypothesis that endodontic instruments
manufactured with the NiTi heat treated M-Wire are more resistant to corrosion in a
solution of NaOCl was refused.
Description
Citation
Collections
Endorsement
Review
Supplemented By
Referenced By
Rights and licensing
Acesso Aberto