Da amplitude à variabilidade: uma abordagem integrada sobre os efeitos da colocação de eletrodos no registro do potencial motor evocado
Loading...
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Universidade Federal de Juiz de Fora (UFJF)
DOI
Abstract
Objective: To investigate the influence of three different electrode montages
on the variability of Motor Evoked Potentials (MEPs) elicited by Transcranial Magnetic
Stimulation (TMS) and recorded using surface Electromyography (sEMG). Methods:
The study was conducted in sequential stages, including a literature review, pilot
studies, and a main experiment. In the main experiment, the right Biceps Brachii
muscle of eight (8) adult participants was assessed using sEMG to record MEPs
obtained from three different electrode montages: (1) bipolar montage with electrodes
positioned over the muscle belly; (2) bipolar montage recommended for MEP recording
in clinical protocols; and (3) pseudomonopolar montage (belly–tendon). The montages
differed in electrode location and interelectrode distance, resulting in different detection
volumes. MEPs were recorded simultaneously across the three configurations.
Results: The bipolar montage (2) demonstrated lower variability compared to the other
two configurations evaluated. Conclusion: Montage (2) provides MEPs with lower
coefficients of variation and higher amplitudes. Significance: Electrode configuration
directly influences MEP amplitude and variability and should be carefully considered
in TMS assessments to ensure greater reliability of measurements.
Description
Citation
Endorsement
Review
Supplemented By
Referenced By
Creative Commons license
Except where otherwise noted, this item's license is described as Acesso Aberto